JG-9015 Salt Spray Test Chamber
JG-9015 Salt Spray Test Chamber Salt Spray Test Chamber Environmental Test Machine Malaysia, Negeri Sembilan (NS), Seremban Supplier, Suppliers, Supply, Supplies | JNG Industry Sdn Bhd
Equipment Usage:
  • The JG-9015 series intelligent salt spray box has highly automated and high-precision test methods. Compared with the ordinary button type salt spray box, it can not only save a lot of management time, but also minimize the test error caused by human factors. Innovative man-machine interaction management mode, accurate test temperature area setting, intelligent automatic water replenishment function, hope to get your approval.


Technical Specifications:

 Model JG-9015-60 JG-9015-90 JG-9015-120 JG-9015-160 JG-9015-200
Studio (L*W*H)cm 60 x 45 x 40 90 x 60 x 50 120 x 80 x 50 160 x 90 x 60 200 x 100 x 60
Volume of outer box 110 x 67 x 118 141 x 90  120 175 x 115 x 125 230 x 125 x 145 280 x 135 x 145
Test room volume 108L 270L 480L 864L 1200L
Volume of brine box 13L 18L 25L 35L 50L

Power supply
 
Single Phase
AC220V 50Hz
2.0KW
Single Phase
AC220V 50Hz
2.5KW
Single Phase
AC220V 50Hz
3.5KW
Three Phase
AC380V 50Hz
5.0KW
Three Phase
AC380V50Hz
7.0KW
Technical parameter Test room temperature RT+5¡æ~50¡æ (Set according to the test standard)
Temperature fluctuation ≤±1.0¡æ
Salt mist settlement  1¡«2ml/h/80cm2
Spray Spray tower  The characteristic cone type disperser adjusts the size of the spray and makes the mist fogging averages.
Injector PYREX Precision nozzle of glass
Spray mode
 
Neutral, acid and copper accelerated salt spray test types are selected, and custom type
can also be chosen, but the parameters must conform to the relevant standard requirements
of salt spray test. In the program test, more test types can be defined.
Mist removal
 
The mist is discharged by air extrusion, and the box lid is conical designed to form a top corner of 120 degrees, which is favorable
for the droplet produced by salt spray not directly dropping on the sample, which ensures the accuracy of the test.
Intelligent touch screen controller Test room temperature control: RT + 5 C ~50 C, precision 0.1 C (double digital display) system VT4810 PID
automatic calculation. 
Liquid expansion super temperature protection controller 0~110 C (Korea)
Saturated barrel temperature controller: 47~70 degrees, 0~1 degrees of accuracy (even display) PID automatic
calculation. 
Time total time: 0~99999 min gap timing: 0~99999 min.
Pressure adjustment: a pressure regulating valve (coarse adjustment, on the back), two pressure regulating valve
(fine adjustment, on the front)
Material Science PVC new class A, transparent observation window 
Experimental Environment 1. Temperature: in order to give full play to the function and performance of the laboratory, the place where the annual temperature is 5 ~35 C should be selected.
2. Relative humidity: no more than 85% R H.3 atmospheric pressure: 86~106kPa                                    
3.There is no strong vibration around
4. Direct radiation without direct sunlight or other heat sources. 
5. There is no strong air flow around it. When the ambient air needs to flow strongly, the air flow should not be blown directly into the box.
6. There is no strong electromagnetic field around it. 
7. No high concentration of dust and corrosive substances around.
Note Special requirements for non-standard, can be customized

Please compare the table carefully so that the model needs to be selected correctly

Please leave your enquiry here, we will reply as soon as possible.


*only support gif, jpeg, jpg, png, pdf